5 ocp calibration, Ocp calibration – B&K Precision XLN60026 - Manual User Manual

Page 200

background image

104

MID2

2.08

1.872 ~ 2.288

MID3

3.468

3.1212 ~ 3.8148

Hi

4.94

4..446 ~ 5.434

XLN15010

Function

Setting

Range of transformation value

Lo

0.5408

0.04 ~ 0.5949

MID1

1.3697

1.2327 ~ 1.5067

MID2

4.16

3.744 ~ 4.576

MID3

6.9368

6.2431 ~ 7.6305

Hi

9.88

8.892 ~ 10.868

6.

If the values are not consistent with your measurements, there

might be issues with the hardware and will need to be inspected
by a trained technician.

7.

Once calibration is finished, the values will be stored into the

internal flash memory and return to the calibration menu.

5.5

OCP Calibration

Follow the step by step instructions below for OCP calibration.

1.

Press

in the calibration menu to enter OCP calibration

menu.

2.

Press

to access the OCP calibration procedure for “Lo”.