B&K Precision XLN60026 - Manual User Manual
Page 26

19
CURR SLEW RATE:
Current ascending/descending slope
(XLN 3640:0.01 - 2.5A/ms)
(XLN 6024:0.01 - 1.2A/ms)
(XLN 8018:0.01 - 0.72A/ms
(XLN 10014:0.01 - 0.48A/ms)
CONNECTOR DROP:
Turns on/off the connector drop calibration
function
EXT FULL VOLT:
External voltage control full-scale setting.
Select between 10 V or 5 V for full-scale
control.
3. PROTECTION SETTING (PROTECTION)
Press
key in the first screen of “Menu Setting” to enter
PROTECTION menu.
OVP:turns on/off the
overvoltage protection
SET: set up the overvoltage protecting
point.
OCP:turns on/off the
overcurrent protection
SET: set up the overcurrent protecting
point.
OPP:turns on/off the overpower
protection
SET: set up the overpower protecting
point.
F
F
F
F
=
=
CV TO CC:
Enable/disable the protection of the change
OVP = OFF
SET =
38.000 V
OCP = OFF
SET = 42.000 A
OPP = OFF
SET = 1440.000 W