2 voltage calibration, Voltage calibration – B&K Precision XLN60026 - Manual User Manual

Page 194

background image

98

5.2

Voltage Calibration

Follow the step by step instructions below for voltage calibration.

1.

Power off the supply and connect its output terminal to a Digital

Volt Meter (DVM), as shown in the figure below. Then set the
meter for DC voltage measurement.

DVM

DC Voltage measurment

Voltage Calibration

DS30052

+

-

2.

Turn on the power supply’s output.

3.

Press

in the third page of Menu Setting to enter

CALIBRATION page. Users can enter the password and then access to
the calibration procedures.