B&K Precision XLN60026 - Manual User Manual

Page 32

background image

25

MIDH

53V

48 - 58 V

Hi

76V

69 - 83 V

XLN10014

Function

Setting

Value

Range of Read-Back Value

Lo

5V

4.5 - 5.5 V

MIDL

33V

29.7 - 36.3 V

MIDH

66V

59 - 72 V

Hi

95V

85 - 104 V

E.

When pressing ENTER at Hi function, and the calibration data

values will be stored in FLASH and unit will return to calibration

main screen.

A.

Press 2 in the calibration main page to enter OVP calibration

page.

OVP CALIBRATION

L

L

a

=

o

L

=

i

B.

Press ENTER to access OVP calibration procedure at Low function.