Key in “13579” as the password to enter the calibration menu.

5.

Press

to access the Voltage Calibration menu, and the

following parameters will be displayed.

6.

With the CALIB VOLT Lo selected, look at the voltage value

displayed on the DVM, and change the value to match the
measured results. Then press

. Repeat and do the same

for CALIB VOLT MIDL, CALIB MIDH, and CALIB VOLT Hi.
The values should meet within the specified range in the below
tables, depending on the model. If it does not, you may have to
inspect the hardware. Please contact B&K Precision.