B&K Precision XLN60026 - Manual User Manual

Page 199

background image

103

calibration menu.

3.

First, input the resistance of the current shunt in mΩ for current

measurement fixture and press

.

4.

Then, observe the voltage value shown on the DVM, and input it

accordingly for CALIB CURR Lo, then CALIB CURR MID1,
MID2, MID3, and Hi accordingly.

5.

Be sure to check that all values are within the range specified by

the below tables, depending on the model.

XLN60026

Function

Setting

Range of transformation value

Lo

0.1352

0.01 ~ 0.1487

MID1

0.3424

0.3082 ~ 0.3766

MID2

1.04

0.936 ~ 1.144

MID3

1.7342

1.5608 ~ 1.9076

Hi

2.47

2.223 ~ 2.717