B&K Precision XLN60026 - Manual User Manual
Page 186

90
CMC?
Description: Queries the measured/readback current output
value of the selected power supply
Query:
CMC?
Return
String:
1.234 A
The selected power supply has a measured output
current of 1.234 A.
CDVC?
Description: Queries the display voltage and current of the
selected power supply.
Example:
CDVC?
Queries channel 1
Return
String:
100.23,1.567
Display voltage is 100.23 V and current is 1.567 A.
Format:
COUT
Description: Configures/Queries the output state of the selected
power supply.
Example:
COUT ON
Sets the selected power supply’s output to ON.
COUT?
Queries the selected power supply’s output state.
COV
Description: Configures/Queries the overvoltage protection
This manual is related to the following products: