beautypg.com

B&K Precision XLN60026 - Manual User Manual

Page 137

background image

41

EXTERN

ENABLE

Enable

+/Enable -

Output

Output

On/Off

button

Display

NOT ACTIVE

(by Default)

Not Active

On

----

ACTIVE

(A Mode)

Opened

Off

Disabled ENA

Shorted

On/Off

Enabled

----

ACTIVE

(B Mode)

Opened

Off

Disabled

----

Shorted

On

Disabled ----

Note:
When changing from

A MODE to B MODE, change EXTERN ENABLE

to

NOT ACTIVE first to disable analog control, then change to B MODE.

Afterwards, exit the menu to verify that the ENA indicator is no longer
displayed before going back to the menu to reset

EXTERN ENABLE to

ACTIVE.

Shut-Off
Pin 15 can be used to shut off the power supply’s output, controlled by an
input trigger signal. The output shuts off with the falling edge or rising
edge of the trigger. This is selectable from the menu by pressing MENU

6-SPECIAL TEST FUNC  2. EXTERNAL CONTROL. Choose between

OFF/LOW (for falling edge trigger) or ON/LOW (for rising edge trigger).

Falling-edge trigger

EXTERN MONITOR= 10V

EXTERN SHUT-OFF= OFF / LOW

EXTERN ENABLE= ACTIVE