that are connected to the Master. Therefore, in order to correctly configure the
correctly, user must set up the Slaves before
To set a XLN3640/XLN6024/XLN8018/XLN10014 to slave mode, press

the Master is set.

,

, and

in the main screen to enter the series/parallel

setting option. And then select the parallel mode by turning the knob (SCPI
command is “PS:MODE PARALLEL”) and then press

to confirm. It

will continue to the next line for the MASTER/SLAVE selection. Turning the
knob to select SLAVE A (SCPI command is “PS:TYPE SLAVEA”) for the
supply and press

to confirm the setting. Using the same procedure,

setup SLAVE B and C for another two (2) supplies as shown below.

L

L

L

=

= L

L

L

To set the master unit, press

,

, and

in the main screen

to enter the series/parallel setting option. And then use the knob to select the
parallel mode (SCPI command is “PS:MODE PARALLEL”) and press

to confirm. It will continue to the next line for the MASTER/SLAVE selection.
Use knob again to select MASTER (SCPI command is “PS:TYPE MASTER”)
and then press

to confirm. After it is confirmed as the Master the

XLN3640/XLN6024/XLN8018/XLN10014 will start searching for all Slaves,
as shown below:

=