B&K Precision XLN60026 - Manual User Manual
Page 185

89
CPV
Description: Configures/Queries the VSET voltage setting of
the selected power supply.
Example:
CPV 150.68
Sets the selected power supply’s VSET voltage
setting to 150.68 V.
CPV?
Queries the selected power supply’s VSET voltage
setting.
CMV?
Description: Queries the measured/readback voltage output
value of the selected power supply.
Query:
CMV?
Return
String:
120.24 V
The selected power supply has a measured output
voltage of 120.24 V.
CPC
Description: Configures/Queries the ISET current setting of the
selected power supply.
Example:
CPC 3.123
Sets the selected power supply’s ISET current
setting to 3.123 A.
CPC?
Queries the selected power supply’s ISET current
setting.
This manual is related to the following products: