Calibration record – the glp file – YSI 600DW-B Sonde User Manual
Page 83

Sondes
Section
2
CALIBRATION RECORD – THE GLP FILE
When any sensor is calibrated, most 6-series sondes will automatically create a file in sonde memory that
provides details of the calibration coefficients before and after the calibration. The file will have a .glp
extension and will have the Circuit Board Serial # as the default filename. The file can be viewed by
following the path File|Directory|File Details|View from the Main sonde menu.
------------------File-----------------
1-Directory
4-View file
2-Upload
5-Quick view file
3-Quick Upload 6-Delete all files
Select option (0 for previous menu): 1
Filename Samples
1-BRIDGE1.dat 19
2-BRIDGE2.dat 27
3-UPLAKE.dat 33
4-CLRLAKE2.dat 167
5-DWNLAKE2.dat 31
6 - 00003001.
glp
39
-------------File details--------------
1-View file
2-File:00003001.glp
3-Samples: 39
4-Bytes: 790
Pressing 1-View file will show the calibration record for the sonde. An example is shown below:
YSI Incorporated
Drinking Water Monitoring Systems Operations Manual
2-76