There are a variety of probe options for the sondes. The figures below
and to the right illustrate the uses of the common tool for port plug
removal. Note that this tool will also be used to install the various
probes.
If the tool is misplaced or lost, you may use 7/64” and 9/64” hex keys as
substitutes.
OPTIC
PORT PLUG
INSTALLATION
TOOL
DO, COND., &
pH/ORP PORT
PLUGS
INSTALLATION
TOOL
NOTE: You may need pliers to remove the ISE port
plugs, but do not use pliers to tighten the ISE probes.
Hand-tighten
only.
Now refer to the figures below to find the probe locations
in your sonde.
600DW-B SONDE BULKHEAD
ISE PORT PLUG
PLIERS
(SLIP JAWS)
6572 Free Chlorine = 3-pin connector
FREE
CHLORINE
6572
CONDUCTIVITY/
TEMPERATURE
6560
ALL ISE
PROBES
6560 Conductivity/Temperature = 6-pin connector
6561 pH = 4-pin connector
6565 Combo pH/ORP = 4-pin connector
6566 Fouling Resistant pH//ORP = 4 pin connector
6561FG pH (Flat Glass) = 4-pin connector
6565FG pH/ORP (Flat Glass) = 4-pin connector
YSI Incorporated
Drinking Water Monitoring Systems Operations Manual
2-11