beautypg.com

YSI 600DW-B Sonde User Manual

Page 82

background image

Sondes

Section

2

time. After the readings have stabilized, press

Enter to confirm the calibration. Then, as instructed, press

any key to return to the Calibrate

menu.


Select the

3-point option for maximum accuracy over the entire range of 0 to 200 ug/L. As for the 2-point

procedure, one of the standards must be 0 NTU. The procedure for this calibration is the same as for a 2-
point calibration, but the software will prompt you to place the sonde in the additional solution to complete
the 3-point procedure.

For all rhodamine calibration procedures, be certain that the standard and sensor are thermally equilibrated
prior to proceeding with the calibration.



REESTABILISHING DEFAULT CALIBRATION – “UNCAL” COMMAND


If, for any reason, you want to return to the factory default settings (slope and offset) for any parameter,
follow the instructions below:

• Activate any calibration screen (1-point, 2-point, or 3-point) where a numerical entry is requested. In

the example below, a 2-point turbidity calibration was selected.

• Instead of entering a numeric value, type the word “uncal” and press Enter.

• An “uncal” entry at any calibration prompt will reset both slope and offset for this parameter to the

factory default settings.
















---------Turbidity calibration--
-------
1-1 point
2-2 point
3-3 point

Select option (0 for previous
menu): 2
Enter 1st Turbid NTU:uncal

NOTE: When using the 650 MDS as the sonde interface device, the “uncal” operation is performed by
holding the

Escape key down and then pressing the Enter key. See Section 3 of this manual for more

details.




YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-75

This manual is related to the following products: