2 sonde communication problems – YSI 600DW-B Sonde User Manual
Page 241
Troubleshooting
Section
6
The following troubleshooting tables can help you if you encounter problems with software,
communication protocol, or sensor malfunctions other than calibration errors.
6.2 SONDE COMMUNICATION PROBLEMS
SYMPTOM POSSIBLE
CAUSE ACTION
REF
Sonde not powered
Check 12 vdc source
7.2
Cable connection is loose
Check both ends of cable; secure
connectors
Damaged connectors
Check pins at both ends, insure
they are straight, dry and clean.
Cannot communicate
with sonde
Com port not selected
Change to other com port, other
peripheral on the same port
(Internal mouse). Try other PC,
650 display/logger or dumb
terminal
2.4.3
Unmatched baud rate between
host and sonde
Match the baud rate
2.4.3
Host is too slow
Use faster computer
Interface cable failure
Check cable for damage. If
necessary, return for service
7
Internal failure
Return sonde for service
7
Scrambled data