YSI 600DW-B Sonde User Manual
Page 32
Sondes
Section
2
studies,
Auto sleep RS232 should usually be “on”. This is described in detail in Section 2.9, Sonde Menu.
When this setup is verified, press
Esc or 0 to return to the Advanced menu.
Select
3-Sensor from the Advanced menu and make certain that the entries are identical to those shown
below.
------------Advanced sensor-----------
1-TDS constant=0.65
2-Cl2 temp co %/C=2.5
3-Cl2 warm up sec=60
4-( )Wait for Cl2
5-Wipes=1
6-Wipe int=1
7-SDI12-M/wipe=1
8-Turb temp co %/C=0.3
9-(*)Turb spike filter
A-Chl temp co %/C=0
B-( )Chl spike filter
Select option (0 for previous menu):
The display under
3-Sensor may be different from the one shown in the example above, depending on the
sensors that are installed in your unit. For example, if you do not have a chlorophyll probe, the last two
entries (which are relevant only to chlorophyll) will not appear.
When this setup is verified, press
Esc or 0 to return to the Advanced menu. For a detailed explanation of
the choices in the Advanced menu, see
Section 2.9.8, Advanced. Press Esc or 0 to back up to the Main
menu.
------------------Main-----------------
1-Run 5-System
2-Calibrate 6-Report
3-File 7-Sensor
4-Status 8-Advanced
Select option (0 for previous menu):
The sonde software is now set up and ready to calibrate and run.
YSI Incorporated
Drinking Water Monitoring Systems Operations Manual
2-25