beautypg.com

User calibration, Cal test, Menus – Ohaus GT Series Electronic Balance User Manual

Page 77

background image

38

MENUS

User Calibration

User Calibration

User Calibration

User Calibration

User Calibration

User calibration is used when it is desired to calibrate the balance using a mass of
known value. To use this calibration feature, proceed as follows:

1.

.

2.

.

3.

.

4.

. Last calibration mass value is displayed

with the first digit flashing.

5.

to change flashing digit.

6.

to accept and proceed to the next digit.

to backup.

7. Set the number to match the value of the selected calibration mass. The

number entered must be at least 25% of the full span value.

8. After the last digit has been accepted,

.

9.

.

10. Place calibration mass on the platform.

11.

.

. The balance

is now calibrated, remove the calibration mass. The balance is now in a
weighing mode.

Cal Test

Cal Test

Cal Test

Cal Test

Cal Test

Cal Test offers a choice of the span calibration value (1/4, 1/2, 3/4 or full). To ensure
reproducibility, this feature allows a check of a known calibration mass against last
stored calibration information.

1.

.

2.

.

3.

.

4.

, = no mass should be on the platform.

5.

= mass which

must be placed on the platform.