Span calibration printout, Linearity calibration printout, Calibration test printout – Ohaus GT Series Electronic Balance User Manual
Page 68: Operation

29
OPERATION
Printing Data (Cont.)
Printing Data (Cont.)
Printing Data (Cont.)
Printing Data (Cont.)
Printing Data (Cont.)
Span Calibration Printout
Span Calibration Printout
Span Calibration Printout
Span Calibration Printout
Span Calibration Printout
With GLP on, when performing a Span
calibration, a printout is automatically
made after the calibration mass is placed
on the platform and
is pressed.
- - - - - SPAN CAL - - - - - -
4/01/95
12:00:00 PM
Bal Id 1234
Cal:
4000.00g
Old:
4000.00g
Dif:
0.00g
Wt. Ref......................................
ID 2056853
PR 100012
Name........................................
- - - - - END - - - - -
Linearity Calibration Printout
Linearity Calibration Printout
Linearity Calibration Printout
Linearity Calibration Printout
Linearity Calibration Printout
When performing a Linearity calibration
with GLP on, a printout is automatically
made after the calibration mass is placed
on the platform and
is pressed.
- - - - - LIN CAL - - - - - -
4/01/95
12:00:00 PM
Bal Id 1234
Cal:
4000.00g
Old:
3999.94g
Dif:
0.06g
Wt. Ref......................................
ID 2056853
PR 100012
Name........................................
- - - - - END - - - - -
- - - - - CAL TEST - - - - - -
4/01/95
12:00:00 PM
Bal Id 1234
Cal:
4000.00g
Act:
4000.04g
Dif:
0.04g
Wt. Ref......................................
ID 2056853
PR 100012
Name........................................
- - - - - END - - - - -
Calibration Test Printout
Calibration Test Printout
Calibration Test Printout
Calibration Test Printout
Calibration Test Printout
When performing a Calibration Test with
GLP on, a printout is available. When the
display indicates the mass value to be
placed on the platform, the balance the
automatically displays the calibration
weight required.