37

MENUS

Span Calibration

Span Calibration

Span Calibration

Span Calibration

Span Calibration

1.

.

2.

.

3.

, = no mass should be on the platform.

4.

= mass which

must be placed on the platform.

5.

allows the selection of either 25%, 50%, 75%, or 100% of full

span to be used to calibrate the balance. Example shows value of 50% for a
4 kg balance.

6. Place required mass on the platform ,

.

. When mass is displayed with current unit indicator,

and the stability indicator (S) is ON, the balance is recalibrated.

7. Remove the mass from the platform

.

Span

calibration is complete. The balance is now in a weighing mode.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: Do not disturb the balance

Do not disturb the balance

Do not disturb the balance

Do not disturb the balance

Do not disturb the balance

when -C- is displayed.

when -C- is displayed.

when -C- is displayed.

when -C- is displayed.

when -C- is displayed.

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

1.

.

2.

.

3.

.

4.

, = no mass should be on the platform.

5.

= mass which

must be placed on the platform.

6. Place the required mass on the platform,

= next mass to be placed on the platform.

7. Place the required mass on the platform,

.

. When mass is displayed with current unit indicator

and the stability indicator (S) is ON, the balance is recalibrated.

8. Remove the mass from the platform

. Linearity

calibration is complete. The balance is now in a weighing mode.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: Do not disturb the balance

Do not disturb the balance

Do not disturb the balance

Do not disturb the balance

Do not disturb the balance

when -C- is displayed.

when -C- is displayed.

when -C- is displayed.

when -C- is displayed.

when -C- is displayed.