Span calibration, Linearity calibration, Menus – Ohaus GT Series Electronic Balance User Manual
Page 76

37
MENUS
Span Calibration
Span Calibration
Span Calibration
Span Calibration
Span Calibration
1.
.
2.
.
3.
, = no mass should be on the platform.
4.
= mass which
must be placed on the platform.
5.
allows the selection of either 25%, 50%, 75%, or 100% of full
span to be used to calibrate the balance. Example shows value of 50% for a
4 kg balance.
6. Place required mass on the platform ,
.
. When mass is displayed with current unit indicator,
and the stability indicator (S) is ON, the balance is recalibrated.
7. Remove the mass from the platform
.
Span
calibration is complete. The balance is now in a weighing mode.
NOTE:
NOTE:
NOTE:
NOTE:
NOTE: Do not disturb the balance
Do not disturb the balance
Do not disturb the balance
Do not disturb the balance
Do not disturb the balance
when -C- is displayed.
when -C- is displayed.
when -C- is displayed.
when -C- is displayed.
when -C- is displayed.
Linearity Calibration
Linearity Calibration
Linearity Calibration
Linearity Calibration
Linearity Calibration
1.
.
2.
.
3.
.
4.
, = no mass should be on the platform.
5.
= mass which
must be placed on the platform.
6. Place the required mass on the platform,
= next mass to be placed on the platform.
7. Place the required mass on the platform,
.
. When mass is displayed with current unit indicator
and the stability indicator (S) is ON, the balance is recalibrated.
8. Remove the mass from the platform
. Linearity
calibration is complete. The balance is now in a weighing mode.
NOTE:
NOTE:
NOTE:
NOTE:
NOTE: Do not disturb the balance
Do not disturb the balance
Do not disturb the balance
Do not disturb the balance
Do not disturb the balance
when -C- is displayed.
when -C- is displayed.
when -C- is displayed.
when -C- is displayed.
when -C- is displayed.