Technical data, 1 specifications, Admissible ambient conditions – Ohaus MB45 MOISTURE ANALYZER Manual User Manual
Page 64

62
MB45
11. Technical Data
11.1 Specifications
Model MB45
Capacity 45g
Readability 0.001g,
0.01%
Results
Moisture (%), dry content (%), Wt (g), Wt (custom),
regain, %
Temp. Settings
50°C to 200°C
Drying Programs
Standard, fast, ramp, step
Switch-off Criteria
Timed, Auto (30, 60, 90 seconds, auto free), manual
Heat Source
Halogen
Calibration External
calibration mass-20g
Dimensions (DxWxH) (in / cm)
14 x 7.5 x 6 / 35.5 x 19 x 15.2
Pan Size
90 mm. diameter
Adjustable Feet and Level
Yes
RS232 Interface
Yes
Weight (lb / kg)
9.8 / 4.5
Shipping Weight (lb / kg)
14.1 / 6.4
Admissible ambient conditions
Indoor use only
Altitude:
Up
to
2000m
Temperature range:
5 ºC to 40 ºC
Atmospheric
humidity:
Maximum
relative
humidity 80% for temperatures up to 31 °C
decreasing
linearly to 50% relative humidity at 40 °C.
Warm-up
time:
At
least
60
minutes after connecting the instrument to the
power
supply;
when
switched
on
from
standby-mode,
the
instrument
is
ready
for
operation
immediately.
Voltage
fluctuations: Mains supply voltage fluctuations up to –15% +10% of the
nominal
range
Over
voltage
category:
II
Pollution
degree: 2
Power
load: Max.
450
W
during
drying
process
Current consumption:
4 A or 2 A, according to the heating element
Power supply voltage:
100 V – 120 V or 200 V – 240 V, 50/60 Hz
(the
voltage
is
given
by
the
heating
element)
Power line fuse:
1 piece, 5 x 20 mm, T6,3 H 250 V