RAMP PROFILE
80
C
100 C
130 C
STEP 1:
STEP 2:
FINAL:
EXIT WITHOUT SAVING
5 MIN
5 MIN
For STEP 1, using the arrow buttons, define the temperature. Using the
arrow buttons, scroll to time.
Using the arrow buttons, define the time.
Repeat the above procedure for step 2 and FINAL step and press the
Enter button.
6.4 Setting the drying temperature
The drying temperature can be set from 50°C to 200°C. Two of the programs under TEST PARAMETERS -
PROFILE (Standard and Fast) use this temperature setting.
TEST PARAMETERS
TEST ID:
PROFILE:
DRY TEMP:
SWITCH OFF:
RESULT:
CUSTOM:
-DEFAULT-
STANDARD
180C
AUTO 60
MOISTURE%
OFF
Using the arrow buttons, scroll to DRY TEMP and press the Enter
button.
130
C
EXIT WITHOUT SAVING
DRYING TEMP
Using the arrow buttons, enter the desired drying temperature and
press the Enter button.
Temperature limitation
The following protective measures have been implemented
to protect the instrument against overheating:
— The Moisture Analyzer is equipped with an overheating
protection.
— At temperatures above 160C, a time limit becomes
active. The higher the temperature, the shorter the time
until the instrument starts to lower the temperature
(envelope curve, see adjacent illustration).
Note: If you are working at temperatures above 180C, we advise you to wait 2 to 3 minutes with open
instrument lid between individual measurements to ensure good reproducibility and avoid overheating of
the instrument.