5 selection of the drying time – Ohaus MB45 MOISTURE ANALYZER Manual User Manual
Page 54

52
MB45
8.5 Selection of the drying time
The Halogen Moisture Analyzer offers two different types of switch-off criteria. A switch-off criterion is
understood to mean the condition which must be met for the Halogen Moisture Analyzer to switch off
automatically and end the drying.
The first type is a timed switch-off. This is used primarily when the drying process does not attain a
constant final value and the sample continuously loses mass over time through decomposition or the
evolution of difficult to volatilize components.
The second type recognizes the end of drying automatically. The integrated balance continuously
determines the weight loss of the sample during drying. If the weight loss per unit of time is less than the
set level, the drying is stopped and the final result is displayed.
Auto Switch Off
The automatic switch-off criterion is divided into three selectable levels:
— A30: Less than 1mg loss in 30 seconds, used for rapidly drying substances.
— A60: Less than 1mg loss in 60 seconds, used for standard samples.
— A90: Less than 1mg loss in 90 seconds, used for slow drying substances (e.g. plastics).
AFree switch-off criterion
The Auto Free switch-off criterion is based on a user defined mean weight loss per unit of time. As soon
as this drops below the preset value, the measurement is automatically ended.
Manual
With this switch-off criterion, the measurement process continues until you stop it with the STOP button.
The elapsed time is shown in the display.
Timed switch-off
With this switch-off criterion, the measurement lasts until the preset drying time has elapsed.