beautypg.com

Care and maintenance, 1 cleaning interior/exterior components – Ohaus MB45 MOISTURE ANALYZER Manual User Manual

Page 59

background image

MB45

57

10. Care and Maintenance

In this section, you will learn how to keep your Moisture Analyzer in good condition and how to replace
expendable parts.

10.1 Cleaning Interior/Exterior Components

To continue to obtain precise measurements, it is advisable to clean the interior components at regular
intervals. Please note the following instructions for cleaning your instrument.

Disconnect the instrument from the power supply before cleaning.

Open the cover and remove the pan handler,
sample pan, draft shield, and heat shield from the
instrument before cleaning.

Use a lint-free cloth for cleaning.

Clean the exterior of the instrument and the drying
compartment components with a mild cleaning
agent. Although the housing is extremely rugged
and resistant to solvents, never use abrasive
cleaning agents or solvents!

Ensure that no liquid enters the interior of the
instrument.

Replace the components after cleaning.

Draft Shield

Pan Support

Heat Shield

Pan Handler

Sample Pan