beautypg.com

3 setting the drying profile, Drying profile, Test parameters – Ohaus MB45 MOISTURE ANALYZER Manual User Manual

Page 33

background image

MB45

31

6.3 Setting the drying profile

The drying profile contains four settings, Standard, Fast, Ramp and Step. Section 6, paragraph 6.4
describes each of the drying profiles. Review the profile descriptions. Depending upon the sample
material to be analyzed, select the appropriate profile and proceed as follows:

TEST PARAMETERS

TEST ID:

PROFILE:

DRY TEMP:
SWITCH OFF:
RESULT:
CUSTOM:

-DEFAULT-

STANDARD

180C
AUTO 60
MOISTURE%
OFF

Standard Profile
Using the arrow buttons, scroll to PROFILE and press the Enter button.

Using the arrow buttons, scroll to STANDARD and press Enter button.

TEST PARAMETERS

TEST ID:

PROFILE:

DRY TEMP:
SWITCH OFF:
RESULT:
CUSTOM:

-DEFAULT-

STANDARD

180C
AUTO 60
MOISTURE%
OFF

Fast Profile
Using the arrow buttons, scroll to PROFILE and press the Enter button.

DRYING PROFILE

FAST

EXIT WITHOUT SAVING