2.2 Selecting the Location
The Moisture Analyzer should always be used in an environment which is free from excessive air
currents, corrosives, vibration, and temperature or humidity extremes. These factors will affect displayed
weight readings.
DO NOT install the Moisture Analyzer:
• Next to open windows or doors causing drafts
or rapid temperature changes.
• Near air conditioning or heat vents.
• Near vibrating, rotating or reciprocating
equipment.
• Near magnetic fields or equipment that
generate magnetic fields.
• On an unlevel work surface.
• In confined areas, allow sufficient space
around the instrument for ease of operation
and keep away from radiating heat sources.
2.3 Installing the Heat Shield, Draft Shield and Pan Support
Lift the cover straight up and Install the heat shield in the base
of the heating chamber.
Install the draft shield element (only one position possible) on
top of the heat shield.
Install the pan support into postion. Turn the pan support until it
engages. In the locked position, the rear arm of the pan support
points directly towards the rear of the analyzer.
Draft Shield
Pan Support
Heat Shield