beautypg.com

6 result, Auto free switch-off (cont), Auto free – Ohaus MB45 MOISTURE ANALYZER Manual User Manual

Page 38

background image

36

MB45

Auto free switch-off (cont)

Using the arrow buttons, scroll to MG, now enter the weight loss (1 mg
to 10 mg).

AUTO FREE

SWITCH OFF AT LESS THAN

5 MG IN

60

SECONDS

EXIT WITHOUT SAVING

Using the arrow buttons, enter the time (10 seconds to 120 seconds).
Press the Enter button.

6.6 Result

Displayed result offers you the choice of whether your display indicates % moisture, % solids, % regain,
grams or custom unit during the measurement process.

TEST PARAMETERS

TEST ID:
PROFILE:
DRY TEMP:
SWITCH OFF:

RESULT:

CUSTOM:

-DEFAULT-
STANDARD
125C
AFREE

MOISTURE%

OFF

Using the arrow buttons, scroll to RESULT and press the Enter button.

% MOISTURE

EXIT WITHOUT SAVING

TEST PARAMETERS

Using the arrow buttons, scroll to either % MOISTURE, % SOLIDS, %
REGAIN, GRAMS or CUSTOM UNITS and press the Enter button.

NOTES: 1. If you change the displayed result during a test, the final

result and the printed results will be in displayed units.

2.

When CUSTOM is selected, the parameters for custom units

are changed in paragraph 6.7.