4 language setting, 5 audible signal level, Language – Ohaus MB45 MOISTURE ANALYZER Manual User Manual
Page 25: Analyzer setup
MB45
23
5.4 Language setting
For convienence, the Moisture Analyzer is equipped with five languages available on all displays. This
procedure permits the selection of any one language.
Press the Setup button. Using the arrow buttons, select LANGUAGE.
Press the Enter button.
LANGUAGE
TEXT LANGUAGE
ENGLISH
EXIT WITHOUT SAVING
Using the arrow buttons, scroll to the desired language of either English,
Deutsch, Espaniol, Francais or Italiano and press the Enter button.
Note: If you select English (US) as the dialog language, the date format is
changed and appears on all records in the American notation
(month/day/year).
5.5 Audible signal level
Certain processes are signalled by an audio signal (e.g. end of drying, entries, error messages, etc.). In
this menu you can select whether the audio signal should be loud, soft or switched off.
ANALYZER SETUP
LOCK OUT
WEIGHT CAL
TEMP CAL
LANGUAGE
BEEPER
TIME-DATE
Press the Setup button. Using the arrow buttons, scroll to BEEPER.
Press the Enter button.
Using the arrow buttons, scroll to the desired setting of either LOW, HIGH
or OFF and press the Enter button. The display returns to ANALYZER
SETUP menu.