Analyzer setup, 1 lock out – Ohaus MB45 MOISTURE ANALYZER Manual User Manual
Page 22

20
MB45
5. Analyzer Setup
The Analyzer Setup menu is activated by pressing the Setup button. The menu contains: lockout, weight
and temperature calibration, language selection, beeper volume settings, time and date format,
communication settings, print criteria, display contrast and brightness settings, and factory reset. Please
review this entire section and the rest of the manual before making changes to the Analyzer. To gain
access to any item in the menu, you can use the arrow buttons to scroll up and down and press Enter to
highlight your selection. To exit the setup menu, press the Display button to return to a test or Test Menu
to change the current test. A description of each menu item and procedure follows:
5.1 Lock Out
Selecting this entry, then setting Lock Out to ON locks out all changes from the Test Library and Analyzer
Setup. You would use this function after the Analyzer has been put into operation, and you do not want
to change any of the test or analyzer setup parameters. “LOCK OUT” is intended to protect against
accidental menu changes.
Press the Setup button. Using the arrow buttons, select LOCK OUT and
press the Enter button.
LOCK OUT
LOCK OUT WILL LIMIT
ACCESS TO MENUS AND
SAVED SETTINGS / TESTS.
LOCK OUT:
OFF
EXIT WITHOUT SAVING
With LOCK OUT highlighted, use the arrow buttons to select either ON or
OFF and press the Enter button. After selection, the display returns to
ANALYZER SETUP menu.