beautypg.com

8 technical data, 9 application – M&C TechGroup SP2400-H Operator's manual User Manual

Page 9

background image

2-1.1-ME

Gas sampling and gas conditioning technology

9

8 TECHNICAL

DATA

Gas sample probe type

SP2000

SP2000-H

SP2300-H

SP2400-H

Part no.

20S1000

20S2000

20S3000

20S3500

Weather protection shield

no

yes

yes

yes

Protection class terminal box

IP54 EN60529

Material filter housing

Stainless steel SS316/316Ti*

PTFE

Titanium

Sealing material

FPM* /7aT** = PTFE -H320/C** = Graphite

Probe flange sealing material

Novapress

Insitu probe tube / prefilter

Optional

Sample pressure max.

0,4-6 bar* abs., /7aT**= 2 bar abs.,
/HP** = 25 bar abs.

2 bar abs.

0,4-6 bar abs.

Ambient temperature

20 °C to +180 °C

-20 °C to +60 °C*;
/PT100, /Fe-CuNi, /Ni-CrNi** = -20°C to +80°C

Filter chamber volume

120 cm

3

Filter porosity

S-2K150= ceramic*, 2micron, /F-0, 1GF150= glass fibre**, 0,1 micron,
/F-3SS150= stainless steel**, 3 micron, /FW= spun glass**

Thermostat, temperature adjust-
ment

0-180°C* -H320/C**= 0-320°C /PT100** /Fe-CuNi** /Ni-
CrNi**

Ready for operation

after 40min -H320/C** = after 60min

Low temperature alarm contact*

contact rating: 250V, 3A~, 0,25A= alarm point: ∆T 30°C

Sample gas outlet cnnection

1x 1/4“ NPTi* tube connection** ø 6, 8 or 10mm
-H320/C**= tube connection 6mm or 8mm*

Backflush/Test gas connection

1/4“ NPTi* /R**, -H320/C**= tube ø 6 mm

Power supply

230V 50/60 Hz, 800W /115V** = 115V 60Hz, 800W (fuse
protection 10A)

Electrical connections

Terminals max. 4mm

2

, 2 x cable gland PG 13,5

Electrical equipment standard

EN 61010, EN 60519-1

Mounting flange

DN65 PN6, form B >DN or ANSI possible** /HP** = DN50 PN25

Mounting flange material

SS316Ti

PTFE

Titanium

Weight

7 kg*

15,4 kg*

15,4 kg*

14,5kg*

*

= Standard

** = Optional

Differential pressure and T90-time at different flow rates
P and T90 at flow of

100

200

500

1000

1500

3000 (only /HF)

Nl/h

P with new filter element
S-2K 150/F-0,1GF150

0,007 0,011 0,020 0,058 0,135 0,240/0,225 bar

P with new filter element
F-3SS150

0,006 0,012 0,040 0,110 0,215 0,405

bar

T90-time without sample
tube/prefilter

6,0 3,5 1,0 <0,5 <0,5 <0,5

sec.



9 APPLICATION

The M&C gas sample probes type SP2000..., SP2300-H and SP2400-H are used for continuous ex-
traction of gases from dust laden, high-temperature and/or humid processes.