8 technical data, 9 application – M&C TechGroup SP2400-H Operator's manual User Manual
Page 9
2-1.1-ME
Gas sampling and gas conditioning technology
9
8 TECHNICAL
DATA
Gas sample probe type
SP2000
SP2000-H
SP2300-H
SP2400-H
Part no.
20S1000
20S2000
20S3000
20S3500
Weather protection shield
no
yes
yes
yes
Protection class terminal box
IP54 EN60529
Material filter housing
Stainless steel SS316/316Ti*
PTFE
Titanium
Sealing material
FPM* /7aT** = PTFE -H320/C** = Graphite
Probe flange sealing material
Novapress
Insitu probe tube / prefilter
Optional
Sample pressure max.
0,4-6 bar* abs., /7aT**= 2 bar abs.,
/HP** = 25 bar abs.
2 bar abs.
0,4-6 bar abs.
Ambient temperature
20 °C to +180 °C
-20 °C to +60 °C*;
/PT100, /Fe-CuNi, /Ni-CrNi** = -20°C to +80°C
Filter chamber volume
120 cm
3
Filter porosity
S-2K150= ceramic*, 2micron, /F-0, 1GF150= glass fibre**, 0,1 micron,
/F-3SS150= stainless steel**, 3 micron, /FW= spun glass**
Thermostat, temperature adjust-
ment
0-180°C* -H320/C**= 0-320°C /PT100** /Fe-CuNi** /Ni-
CrNi**
Ready for operation
after 40min -H320/C** = after 60min
Low temperature alarm contact*
contact rating: 250V, 3A~, 0,25A= alarm point: ∆T 30°C
Sample gas outlet cnnection
1x 1/4“ NPTi* tube connection** ø 6, 8 or 10mm
-H320/C**= tube connection 6mm or 8mm*
Backflush/Test gas connection
1/4“ NPTi* /R**, -H320/C**= tube ø 6 mm
Power supply
230V 50/60 Hz, 800W /115V** = 115V 60Hz, 800W (fuse
protection 10A)
Electrical connections
Terminals max. 4mm
2
, 2 x cable gland PG 13,5
Electrical equipment standard
EN 61010, EN 60519-1
Mounting flange
DN65 PN6, form B >DN or ANSI possible** /HP** = DN50 PN25
Mounting flange material
SS316Ti
PTFE
Titanium
Weight
7 kg*
15,4 kg*
15,4 kg*
14,5kg*
*
= Standard
** = Optional
Differential pressure and T90-time at different flow rates
P and T90 at flow of
100
200
500
1000
1500
3000 (only /HF)
Nl/h
P with new filter element
S-2K 150/F-0,1GF150
0,007 0,011 0,020 0,058 0,135 0,240/0,225 bar
P with new filter element
F-3SS150
0,006 0,012 0,040 0,110 0,215 0,405
bar
T90-time without sample
tube/prefilter
6,0 3,5 1,0 <0,5 <0,5 <0,5
sec.
9 APPLICATION
The M&C gas sample probes type SP2000..., SP2300-H and SP2400-H are used for continuous ex-
traction of gases from dust laden, high-temperature and/or humid processes.