M&C TechGroup SP2400-H Operator's manual User Manual
Page 17
![background image](/manuals/255963/17/background.png)
2-1.1-ME
Gas sampling and gas conditioning technology
17
Remove the probe's protection shield after opening the two clamping devices;
Turn toggle screw „A“ approx. 1 turn counter-clockwise, that the filter housing lid will lift up;
Turn handle „C“ in position „E“;
Turn clamp „B“ counter-clockwise in direction „G“;
Take toggle screw „A“ and pull out the filter housing lid;
Check tight fit of filter element and tighten it if necessary (hand tight) or change filter element and
corresponding sealings;
W A R N I N G !
When working during operation:
High surface temperatures!
Touching the surfaces can result in burns. Wear protective gloves
and any unauthorized access to the probe must be made impos-
sible !
Push in the filter housing lid into the filter area;
Turn clamp „B“ clockwise and turn with the handle „C“ the ringsscrew „D“ in position „E“, that the
clamp „B“ will latch into the ringscrew „D“ and the threaded bolt „H“. It could be neccessary to move
the filter housing lid a little bit forward and backward;
Turn handle „C“ in position „F“ and fasten the filter housing lid by turning the toggle screw „A“
clockwise by hand.
The following pictures should explain the above mentioned steps.
Figure 4
Removing the new filter housing lid