W A R N I N G !
When working during operation:
High surface temperatures!
Touching the surfaces can result in burns. Wear protective gloves
and any unauthorized access to the probe must be made impossi-
ble !
17.1
CHANGING FILTER ELEMENT AND CHECKING SEALINGS
Close ball valve (if fitted). Flush probe in case of toxic gases!
Remove probe protection shield after opening pressure clamps.
Remove filter holder with lid as described in chapter 14.1.
Screw out the filter's knurled screw I, if existing put adapter L into the new filter element and renew
filter element J.
Check filter element seals K and replace if necessary.
Check O-rings in the lid (flat graphite seal for /320H..., PTFE rings for /7aT) and change
if necessary.
Figure 19
Filter elements and gaskets
Clean filter chamber.
It is now also possible to rod through the probe tube in order to remove deposits. Take care to
avoid breakages when handling probe tubes made of alu-oxide !
Replace filter holder with lid, turn handle C into position E and tighten lid with toggle A again.