Gas sampling and gas conditioning technology

2-1.1-ME

11.2 TEMPERATURE

CONTROLLER

In standard version -H, or with option –H320/C temperature control is ensured by way of the capillary
tube thermostat which is built into the connection box and which has a regulation range between 0°C
to 180°C (-H) or 0°C to 320°C (-H320/C). Rated values can be set to a maximum 180°C resp. 320°C.
The temperature has an excess temperature limiter which permanently switches off the heating if the
set temperature is exceeded by 30°C. The heating can be switched back on again by pressing the
green RESET-button which is located below the aperture in the thermostats mounting.
The thermostat has a temperature failure alarm which activates a contact if the temperature falls 30°C
below the set temperature. This status alarm is located on the terminal strip as a potential-free switch
over contact.

If temperature control is to be achieved using external electronic equipment, a PT100 (a max. of two
can be built in) or an Fe-CuNi resp. Ni-CrNi thermoelement is provided as a temperature sensor.
M&C

also supplies suitable temperature controller, e.g. Type 70304G (see data sheet 2-5.1) which

also can be mounted directly at the gas sample probe (max. ambient temperature +45°C).

11.3

SAMPLE TUBE AND PREFILTER POSSIBILITIES

Depending on the process gas temperature and composition, probe tubes made of different materials
are used with ¾" connections.

Sample tube
Type

Part no.

Temperature
max. °C

Material
tube / connection part

Length

1)

mm

Length
max. mm

Tube ø
o/i „d1“ mm

SP2000/PV 20S9070

90

PVDF/

PTFE-GV

1000 1500 25/21

SP32** 20S9280

90

PVDF/

PTFE-GV 800

800

50/44

SP2000/T 20S9083

160

PTFE/

PTFE-GV 500 500 25/15

SP2000/Ti 20S9075

400

Titanium

1000

2500 25/22

SP210/SS 02S9200

600

SS316Ti

1000

2000 12/10

SP2000/SS 20S9065

600

SS316Ti

1000 2500 25/22

SP2000/SS-Vm 20S9067 600

SS316Ti

1000

2500

25/06

SP2000/HC 20S9090

900

Hastelloy

C4

1000 2500 25/22

SP2000/KA

20S9080

1300

Kanthal / SS316Ti

1000

1500

27/20

SP2000/IN 20S9077

1100

Inconel

1000 2000 25/22

SP2000/CR-2* 20S9098

1400

Cr

AL

2

O

3

/ Hastelloy C4

900

900

22,5/13

SP2000/CR-20* 20S9099 1400

Cr

AL

2

O

3

/ Hastelloy C4

1200

1200

22,5/13

SP2000/AO + ↓ 20S9385 1800

Aluminiumoxide

2)

/

1000

1500

24/18

Adapter 20S9395

(600) /

SS316Ti

Adapter

20S9397

(900)

/ Hastelloy C4

* Sample tube with support adapter. Max.Temperature concern the area of the support adapter on approx. 200 mm length depend on material 600/900 °C.
** For gas sampling downstream wet scrubbers for droplet separation

1)

standard

2)

Please pay attention to the characteristic feature of ceramic in case of high and changing temperatures!

Other materials or executions on request.


For more information about sample tubes see data sheet 2-1.1.0.6

To prevent condensation between sample point and heated gas sample probe or in case of conden-
sation in the area of the connection piece heated sample tubes made of stainless steel with flange
connection DN65 PN6 are used.