KLP 073008

B-11

FIGURE B-4. PROGRAMMING AS A CURRENT STABILIZER

B.43 [SOURce:]CURRent:[:LEVEL]TRIGgered[:AMPlitude] COMMAND

CURR:TRIG

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]
= digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1

Description: Programs current value to be transferred to output by *TRG commands. Actual output current

will depend on load conditions. If the value exceeds the maximum for the model being programmed,
error message -222,”Data out of range” is posted in output queue. If value exceeds CURR:LIM:HIGH
value, a value corresponding to the current limit will be programmed. NOTE: A voltage trigger level
must be entered at least once. Related Commands: CURR. (See example, Figure B-1.)

B.44 [SOURce:]CURRent:[:LEVEL]TRIGgered[:AMPlitude]?

QUERY

CURR:TRIG?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]CURRent[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: = digits with decimal point and Exponent, e.g., 2.71E+1 for 27.1

B.45 [SOURce:]FUNCtion:MODE COMMAND

FUNC:MODE

Syntax:

Short Form:FUNC:MODE {VOLT | CURR}
Long Form: FUNCtion:MODE {VOLT | CURR}

Description: Establishes the operating mode of the power supply. VOLT = Constant Voltage mode (CV).

CURR = Constant Current mode (CC). Default operating mode is CV.

NOTE: The following example assumes KLP 75-33-1200 with virtual model set for 36V, 33.33A and output

operating in current stabilization mode. See Figure B-3 for directions on programming virtual
model limits.

VOLT 32.1;CURR 4

Programs output voltage for 32.1V, output current for 4A.

OUTP ON

Enables output. NOTE: Load determines output operating mode; if

load resistance is less than 32.1/4, output will operate as current sta-
bilizer.

MEAS:CURR?

Returns measured output current, approximately 4E0 assuming cur-

rent stabilizer operation.

CURR?

Returns 4E0 (4A, programmed output current).

CURR 3.3E-1

Programs output current to 0.33A. NOTE: For this model, 0.4A is the

minimum load current (see Table 1-1), therefore although 0.33 value
is programmed; load current will not be less than 0.4A. No explicit
indication is given regarding this change.

CURR?

Returns 4E-1 (0.4A).

CURR? MAX

Returns 3.333E1 (33.33A, the virtual model current setting).

CURR:PROT .5

For this model, minimum overcurrent protection is 0.6 x 33.33 =

20A,(see PAR. 3.2.7), therefore power supply returns error message
-222 (Data out of range, ESR bit 4 =1) and the limit remains
unchanged.

CURR:PROT 25

Programs overcurrent protection limit to 25A.

CURR:PROT?

Returns 2.5E1 (25A).

CURR 26

Returns error message -301 (Value bigger than limit, ESR bit 3=1) and

current setting remains unchanged at 0.4A.

CURR:PROT?MAX

Returns 4E1 (maximum allowable overcurrent protection limit for

model is 40A, 20% above maximum current (33.33A) for KLP 75-33-
1200, regardless of virtual model setting).