B-22

KLP 073008

-221,”Settings conflict”

Execution error bit 4

E.g. , Invalid password from syst:pass:cen command. or Calibra-
tion state not enabled but CALibrate command received.

-222,“Current, Voltage or Data out of
range”

Execution error bit 4

Value (current or voltage) exceeds power supply rating or (data)
exceeds acceptable command parameters.

-223,”Data format error”

Execution error bit 4

Multiple decimals in digit, Multiple E, etc.

-224,“Illegal parameter value”

Execution error bit 4

For example, OUTP 2 instead of OUTP 1.

-226,“Lists not same length”

Execution error bit 4

During a LIST command, number of DWEL list entries was not
equal to 1 and did not match number of LIST:VOLT or LIST:CURR
entries.

-240,“Hardware error”

Execution error bit 4

Power supply did not respond to command.

-241,”Hardware missing”

Execution error bit 4

Requesting device 2 status (INST:NSEL 2).

-280, “Program Error“

Execution error bit 4

LIST execution error, e.g., program already running.

-282,”Illegal program name”

Execution error bit 4

Executing prog:run without LIST:SEQ command.

-301,”Value bigger than limit”

Device Error bit 3

E.g., requesting a voltage or current exceeds the limit.

-310,”System Failure”

Device Error bit 3

Program checksum incorrect.

-311,”Memory Error”

Device Error bit 3

E.g., power-up NV RAM error, or CALibrate:STORe error.

-313,”Calibration Memory Lost”

Device Error bit 3

No valid calibration found.

-314,”Save/recall memory error”

Device Error bit 3

Using cell other than 1 - 41 for SAV and RCL.

-315,”Configuration Memory Error”

Device Error bit 3

Configuration Memory Error.

-330,”Self Test Failed”

Device Error bit 3

Self test failed.

-340,”Calibration Failure”

Device Error bit 3

Unable to perform auto calibration.

-341,”Non Volatile Mem. CRC error”

Device Error bit 3

Power supply constants may be corrupted; recalibration may be
necessary.

-350,”Queue Overflow”

Device Error bit 3

More than 15 errors are in queue. Error queue was full, error
events have been lost.

-400,”QueryError”

Query Error bit 2

Data requested was lost due to 253 character limit or KLP output
buffer.

-410,”Query interrupted

Query Error bit 2

New command sent before data from previous query read. Previ-
ous query data lost.

-420,”Query Unterminated”

Query Error bit 2

Controller sent query but did not accept data from power supply.
See VOLT? (PAR. B.48), CAL:SAVE (PAR. B.4) commands.

-430,Query Deadlocked

Query Error bit 2

Over 255 characters received in single input string.

TABLE B-4. ERROR MESSAGES (CONTINUED)

ERROR MESSAGE

ESR ERROR BIT SET

(SEE PAR. A.5)

EXPLANATION

This manual is related to the following products: