Page 100: Source, B.57, Statu, Volt:trig, Stat:oper:cond
![background image](https://www.manualsdir.com/files/860634/content/doc100.png)
B-14
KLP 073008
FIGURE B-5. PROGRAMMING AS VOLTAGE STABILIZER
B.56 [SOURce:]VOLTage:[:LEVel]TRIGgered[:AMPlitude]? QUERY
VOLT:TRIG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value:
Description: Returns value representing voltage value to be programmed by *TRG command established by
VOLT:TRIG command). (See example, Figure B-1.)
B.57 STATus:OPERation:CONDition QUERY
STAT:OPER:COND?
Syntax:
Short Form: STAT:OPER:COND?
Long Form: STATus:OPERation:CONDition?
Return Value:
Description: Returns the value of the Operation Condition Register (see Table B-2). The Operation Condition
Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-6.)
NOTE: The following example assumes KLP 75-33-1200 with virtual model set for 36V, 33.33A and
output operating in voltage stabilization mode. See Figure B-3 for directions on programming
virtual model limits.
VOLT 21.8;CURR .5
Programs output voltage for 21.8V, output current for 0.5A.
OUTP ON
Enables output. NOTE: Load determines output operating mode; if
load resistance is greater than 21.8/0.5, output will operate as
voltage stabilizer.
MEAS:VOLT?
Returns measured output voltage, approximately 2.18E1 (21.8V)
assuming voltage stabilizer operation.
VOLT?
Returns 2.18E1 ( (21.8V, programmed output voltage).
VOLT 2.15
Programs output voltage to 2.15V.
VOLT?
Returns 2.15E0 (2.15V, programmed output voltage).
VOLT? MAX
Returns 3.6E1 (36V, the virtual model voltage setting).
VOLT? MIN
Returns 0 (minimum allowable program voltage).
VOLT:PROT 6.5
For this model, minimum overvoltage protection is 0.2 x 75 = 15V,
therefore power supply returns error message -222 (Data out of
range, ESR bit 4 =1) and the limit remains unchanged.
VOLT:PROT 25
Sets overvoltage protection limit to 25V.
VOLT:PROT?
Returns 2.5E1 (25V).
VOLT 26
Returns error message -301 (Value bigger than limit, ESR bit 3=1)
and current setting remains unchanged at 2.15V.
VOLT:PROT?MAX
Returns 9E1 (maximum allowable overvoltage protection limit for
model is 90V, 20% above maximum voltage (75V) for KLP 75-33-
1200, regardless of virtual model setting).
TABLE B-2. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,
AND OPERATION EVENT REGISTER BITS
CONDITION
NU
PR
NU
CC
NU
CV
NU
WTG
NU
CAL
BIT
15
14
11-13
10
9
8
7 - 6
5
4 - 1
0
VALUE
32,768 16,384
2048 -
8192
1024 512
256
128
- 64
32
16 -
2
1
CAL - INTERFACE COMPUTING NEW CALIBRATION
CONSTANTS
CC -
POWER SUPPLY IN CONSTANT CURRENT
MODE
CV -
POWER SUPPLY IN CONSTANT VOLTAGE
MODE
NU -
NOT USED
PR -
PROGRAM RUNNING
WTG - WAIT FOR TRIGGER (SET BY TRIG
SUBSYSTEM)