B.84 system:language? query, B.85 system:password:cenable command, B.86 system:password:cdisable command – KEPCO KLP Series (older -1200 models) Operator Manual User Manual
Page 106: B.87 system:password:new command, B.88 system:password:state? query, B.89 system:security:immediate command, B.90 system:set command, B.84, System:l, B.85
B-20
KLP 073008
been selected at the factory to permit the VISA demonstration program (which uses older command
formats) to operate. SYST:LANG SCPI disables responses to older command formats.
B.84 SYSTem:LANGuage? QUERY
SYST:LANG?
Syntax:
Short Form: SYST:LANG?
Long Form: SYSTem:LANGuage?
Return Value:
Description: Identifies whether unit responds to older command formats. SCPI means the unit will not respond
to older formats such as LIST:TIME. COMP means the unit will respond to older formats such as
LIST:TIME for compatible operation with software written for KLP power supplies. up to Revision 1
B.85 SYSTem:PASSword:CENable COMMAND
SYST:PASS:CEN
Syntax:
Short Form: SYST:PASS:CEN
Long Form: SYSTem:PASSword:CENable
Description: Sets the password enable state if the value matches the current password. This command allows
other commands such as the CAL commands to operate.
B.86 SYSTem:PASSword:CDISable COMMAND
SYST:PASS:CDIS
Syntax:
Short Form: SYST:PASS:CDIS
Long Form: SYSTem:PASSword:CDISable
Description: Clears the password enable state if the value matches the current password.
B.87 SYSTem:PASSword:NEW COMMAND
SYST:PASS:NEW
Syntax:
Short Form: SYST:PASS:NEW
Long Form: SYSTem:PASSword:NEW
Description: Establishes new password. The old (current) password is checked, then replaced by the new pass-
word.
B.88 SYSTem:PASSword:STATe? QUERY
SYST:PASS:STAT?
Syntax:
Short Form: SYST:PASS:STAT?
Long Form: SYSTem:PASSword:STATe?
Return Value:
Description: Returns a 1 if the password state is enabled or a 0 if it is disabled.
B.89 SYSTem:SECurity:IMMediate COMMAND
SYST:SEC:IMM
Syntax:
Short Form: SYST:SEC:IMM
Long Form: SYSTem:SECurity:IMMediate
Description: Initializes all NVRAM variable to factory defaults. This includes default virtual model, corresponding
overvoltage and overcurrent limits; all memory locations are initialized. NOTE: System password must
be enabled prior to issuing this command (see PAR. B.85).
B.90 SYSTem:SET COMMAND
SYST:SET
Syntax:
Short Form: SYSTem:SET {CM0 | DC0 | DC1 | LF0 | LF1 | KL0 | KL1 | L100 | M40 |}
Long Form: SYSTem:SET {CM0 | DC0 | DC1 | LF0 | LF1 | KL0 | KL1 | L100 | M40 |}
Description: Establishes Device Clear, Line Feed, and Reset functions. Sending SYST:SEC:IMM sets LF1,
DC0, and RO0 (as if CM0 was sent). Sending SYST:SET CM1 sets the KLP to operate in compatible
mode and have all GPIB functions compatible with software version 2.9 and lower units. Sending
SYST:SET CM0 sets the unit to be fully SCPI 1997 compliant.
CM0
Establishes DC0, LF0, conditions described below. (SCPI 1997 Standard compliance)
DC0
Device Clear functions per IEEE 488.2 Standard. (No effect on the device (power supply),
only clears internal status registers.)
LF0
Line Feed not provided upon empty buffer condition
DC1
Device Clear functions identical to *RST (Output set to 0V, voltage mode and output set to
OFF except if RO1 (see below) is set.)
LF1
Line Feed provided if buffer is empty and a read is performed.
KL0
Enables front panel controls
KL1
Disables front panel controls.
L100
Number (100) of locations programmable using LIST commands.