4 quick boot, 5 analog input full scale calibration, 6 calibration – KEPCO KLP Series (older -1200 models) Operator Manual User Manual
Page 52: Quick boot -12, Analog input full scale calibration -12, Calibration -12, R. 3.2.11.4, R. 3.2.11.5), R. 3.2.11.6), 4 if
3-12
KLP073008
NOTE:
Once the relay turns on in V->C or C->V mode, the relay will go off again when the
unit transitions back to the original mode, or when the output is turned off.
4. When the relay control mode is set to MAN, the DC VOLTS display shows whether the relay
is ON or OFF. To change the state of the relay press in and rotate either VOLTAGE or CUR-
RENT control until the desired relay state is displayed, then release the control. Press the
DC OUTPUT switch to immediately set the relay to the selected state.
5. Press the DC OUTPUT switch to exit relay control and advance to the next item in the UTIL
menu.
NOTE: If a SYSTem:SECurity:IMMediate command is issued, the above relay control settings
are reset to the factory default: FLT - NORM.
3.2.11.4
QUICK BOOT
Enter UTIL menu and rotate either VOLTAGE or CURRENT control until status display reads
QkBT - Quick Boot (see PAR. 3.2.11). If Quick boot is ON, then the next time the unit is powered
up it will go directly to setpoint mode (status display reads SET, see PAR. 3.2.6) without display-
ing MODL,VIRT, and PROT settings.
• The DC VOLTS display indicates whether quick boot is ON or OFF.
1. Press in and turn either VOLTAGE or CURRENT control until the desired quick boot state is
displayed,
2. Press DC OUTPUT switch to save. The program then advances to the next item in the UTIL
menu.
3.2.11.5
ANALOG INPUT FULL SCALE CALIBRATION
Enter UTIL menu and rotate either VOLTAGE or CURRENT control until status display reads
INVC (see PAR. 3.2.11).
• The DC VOLTS display shows the analog input voltage value that produces the full
scale output voltage established by the virtual model.
• The DC AMPS display shows the analog input voltage value that produces the full scale
output current established by the virtual model.
1. To change the settings tap the DC OUTPUT switch.
2. Rotate VOLTAGE control to set the analog voltage value required to produce the full scale
output voltage (for fine adjustment press in VOLTAGE control and hold while rotating).
3. Rotate the CURRENT control to set the analog voltage value required to produce the full
scale output current (for fine adjustment press in CURRENT control and hold while rotating).
4. Tap the DC OUTPUT switch to accept the new settings or use a thin tool (e.g., a paper clip)
to press the FUNCTION switch to cancel the operation.
3.2.11.6
CALIBRATION
Enter UTIL menu and rotate either VOLTAGE or CURRENT control until status display reads
CAL (see PAR. 3.2.11).