Table 2-2. analog i/o switch functions, Analog i/o switch functions -3, Ieee 488 port connector (j4) pin assignments -3 – KEPCO KLP Series (older -1200 models) Operator Manual User Manual

Page 29

background image

KLP-HV 073008

2-3

TABLE 2-2. ANALOG I/O SWITCH FUNCTIONS

DIP SWITCH

POSITION

FUNCTION

SETTINGS

1

Allows analog programming commands for voltage to be provided
by either variable voltage (PAR. 3.7.3) or resistance (PAR. 3.7.2).
When resistance mode is selected, an internal 1mA current genera-
tor provides the reference for voltage programming. (PAR. 3.7.2)

ON:

Selects Resistance

OFF: Selects Voltage (factory

default)

2

Allows analog programming commands for current to be provided
by either variable voltage (PAR. 3.7.3) or resistance (PAR. 3.7.2).
When resistance mode is selected, an internal 1mA current genera-
tor provides the reference for current programming.

ON: Selects

Resistance

OFF: Selects Voltage (factory

default)

3

When enabled (ON), allows retention of the last entered setpoint val-
ues for voltage and current in non-volatile memory for recall at the
next power-up sequence. (see PAR. 3.2.6.3). If not selected (OFF,
default condition), the stored values will be 0 volts, minimum amps.
The switch setting is detected upon unit power-up.

ON:

Save and recall previous set-
point values upon power-up.

OFF: Power-up set to 0 Volts and

minimum current (factory
default).

4

NOT USED

5

Allows locking of the front panel controls. (see PAR. 3.2.2.1)

ON:

Local controls locked.

OFF: Local controls enabled (fac-

tory default).

TABLE 2-3. IEEE 488 PORT CONNECTOR (J4) PIN ASSIGNMENTS

CONNECTOR

PIN

SIGNAL NAME

FUNCTION

IEEE 488

PORT

1

D

I

01

I/O Line

2

D

I

02

I/O Line

3

D

I

03

I/O Line

4

D

I

04

I/O Line

5

EOI

End or Identify

6

DAV

Data Valid

7

NRFD

Not Ready for Data

8

NDAC

Not Data Accepted

9

IFC

Interface Clear

10

SRQ

Service Request

11

ATN

Attention

12

SHIELD

Shield

13

D

I

05

I/O Line

14

D

I

06

I/O Line

15

D

I

07

I/O Line

16

D

I

08

I/O Line

17

REN

Remote Enable

18

GND

Ground (signal common)

19

GND

Ground (signal common)

20

GND

Ground (signal common)

21

GND

Ground (signal common)

22

GND

Ground (signal common)

23

GND

Ground (signal common)

24

LOGIC GND

Logic Ground

This manual is related to the following products: