Precision Digital PD6310 User Manual
Page 85
Model PD6210 & PD6310 Batch Controllers
Instruction Manual
85
Controller Copy or Cloning Instructions
!
Caution!
Do not connect the two controllers to the same signal
source while cloning. Internal calibration may be
affected.
1. Connect two controllers using a PDA1200 controller copy cable.
Warning!
Use PDA1200 controller cable only!
Using standard CAT5 or other cable will cause
damage to both controllers.
2. Do not connect the two controllers to the same signal source.
3. Power up both controllers. Leave clone controller in Run Mode.
4. Enter
the
Advanced Features
menu of the Master controller see
ADVANCED OPERATION AND PROGRAMMING page 69.
5. Scroll
to
the
Copy
function using the Up arrow button then press
Enter.
6. The controller displays the message Send. Press Enter, the display
flashes while sending data. The message done is displayed when
copying is completed.
Send
Copy
done
Copy
Display flashes
while sending
data
Copy
7. The Clone controller displays the message CPY rC while being pro-
grammed then the message done when copying is completed. The
controller initializes and returns to Run Mode using the same set-
tings as the Master.
8. If
controller
to be cloned does not respond to the data being sent,
refer to Copy Function Requirements above.