beautypg.com

Powered leakage tests, Your hal tester – Seaward ClareHAL User Manual

Page 96

background image

95

Powered Leakage Tests

Power Output Rating
Test Voltage

110V – 230V AC 20A nominal (Vin-
Vout)

Maximum Power Output

Up to 5.0kVA (dependant on mains
supply rating)



Output Power Measurement

Single Phase Power Measurement
Display Range and Accuracy

0.02kVA-5.00kVA ±2% ±0.02kVA

Selectable Range of Pass/Fail Levels 0.001kVA-5.000kVA

Single Phase Power Factor Measurement

Display Range and Accuracy

Ratio 0.000 -1.000 ± 0.030

Selectable Range of Pass/Fail Levels 0.000-1.000


Leakage Current Measurement

Display Range and Accuracy

0.10mA-20.00 mA ± 1% ±5 Counts

Display

Resolution

0.01mA

Selectable Range of Pass/Fail Levels 0.01mA-20mA

Touch Current Measurement
Display Range and Accuracy

0.02mA - 5.00 mA ± 1% ±5 Counts

Display

Resolution

0.01

Selectable Range of Pass/Fail Levels 0.02mA-5.00mA

Measuring

Device

IEC/EN 60990 Fig 4

(IEC/EN 60990 Fig 3 or Fig 5 Upon Request)

Your HAL tester