Your hal tester – Seaward ClareHAL User Manual

Page 159

background image

158

For PWR tests this field is reserved and should be set to zero.

time_hold`

For HV and EBOND tests this is the length of the hold phase in tenths of
a second.
For PWR tests this is the duration for which power is applied in tenths of
a second.

time_ramp_zero

For HV and EBOND tests this is the time to ramp down to zero output in
tenths of a second.
For PWR tests this field is reserved and should be set to zero.

lo_limit1

The primary lower limit for all test types. This lower limit is only checked
after the end of a test.

hi_limit1

The primary upper limit for all test types. For HIPOT and PWR tests the
test will fail immediately if this upper limit is exceeded. For other test
types this limit is only checked after the end of a test.

lo_limit2

Required only for PWR test types this field defines the lower kVA limit.
This limit is only checked after the end of a test.

hi_limit2

Required only for PWR test types this field defines the upper kVA limit.
This limit is only checked after the end of a test.

lo_limit3

Required only for PWR test types this field defines the lower power
factor limit. This limit is only checked after the end of a test.

hi_limit3

Required only for PWR test types this field defines the upper power
factor limit. This limit is only checked after the end of a test.