Your hal tester – Seaward ClareHAL User Manual

Page 35

background image

34

For all units: -

The last icon denotes the S

Start conditions for the test. They are as Fig.


1. Reset E

External safety guard switch to ‘Closed’, followed by

pressing G

Green to start each repeat of the test. (Start Ext+Green)

2. Reset E

External safety guard switch to ‘Closed’, followed by

pressing Green once to start the test (1

st

loop). N

No subsequent

pressing of Green button needed for any repeat tests. Repeat
loops will immediately convene whilst guard switch is closed.
(Start Ext+Green) for 1

st

loop & (Start None) on any repeats.

3. Reset

E

External safety guard switch to ‘Closed’ to start test and

each repeat of the test. (Start Ext)

4. Reset E

External safety guard switch to ‘Closed’ to start the test

cycle. N

No subsequent stimulus needed for repeat tests. Repeats

will automatically be performed, whilst the safety guard is closed.
(Start Ext) for 1

st

loop & (Start None) on any repeats

5. Press G

Green to start test and each repeat of the test. External

safety guard switch must remain closed for test to begin. (Start
Green)

6. Press

G

Green once to start test. N

No subsequent stimulus needed for

any repeat loop tests. Repeat loop tests will automatically start.
External safety guard switch must remain closed for test to begin.
(Start Green) for 1

st

loop & (Start None) on any repeats.

7. N

No start stimulus required. The test begins immediately after the
product data has been entered. This mode is usually a follow-on
test from a previous test.

Use this with caution

. (Start None) for all

tests

Fig. 1 Fig. 2 Fig. 3 Fig. 4 Fig. 5 Fig. 6 Fig.7

Your HAL tester