Performing a manual test, Your hal tester – Seaward ClareHAL User Manual

Page 53

background image

52

Performing a Manual Test

After selecting the manual test display, you can set test parameters.

The user can change an individual field setting when the field is shown
highlighted (use encoder to highlight any fields to be changed). Then
press R

Rotate Field fast key (highlighted when selected), and use the

encoder (to cycle through) to change the variable in the selected field.
Before moving to next field you must deselect Rotate Field. Repeat
process until all desired field settings are selected as required.

Now these manual test settings can also be saved to a list of file names
(max. of ten) which can be recalled for future tests. Press S

Save to save

current setting to highlighted file or select a previously saved file and
press R

Recall will bring those settings to the manual test display screen.

Use Edit Name key and the encoder to edit any file name if necessary.

To start the test, select (highlight) the ‘START’ field, close the safety
guard and then press G

Green. Pressing Red or Abort Test fast key will

abort the test at any time.

If User wishes to print out the test results immediately after the test,
pre-select the After Test: D

Download option at System Configuration

setup. Alternatively any stored test results can be printed through the
Download Results option at Main menu.

Note

In manual mode you will not get Abort Options, when Abort
stimulus is initiated the test will simply stop and revert to the
Meter display. The aborted test will not be stored.

Your HAL tester