Earth / ground bond test, Insulation test, Your hal tester – Seaward ClareHAL User Manual
Page 94

93
Earth / Ground Bond Test
Test Voltage
Nominal 6V AC
Frequency Independent of Supply
50 or 60 Hz)
Display Range and Accuracy
0-1500m ±2% ±5 Counts
Display
Resolution
1
m
Compliance Test Ranges
Current/Load Resistance/Time
5 A - 1000m - Continuous
10 A - 500m - Continuous
25 A - 200 m - Continuous
30
A
-
150
m - 60 secs
40A - 100 m - 60 secs
Settable Output Current Range
0.1A – 40.0A
Selectable Pass/Fail Levels
0-1500 m
Remote Test Initiation – via No-Burn Ground Bond Probe
Insulation Test
DC Output Voltage
250V, 500, or 1000V Selectable
Display Range
0.01M-500M
Display Range / Accuracy 0.03M - 350 M ± 5% ± 5 Counts
350M -500 M Indication Only
Display Resolution
0.01M
Pass/Fail Level
0.00M-500.0M
Your HAL tester