Digit 10 – arc detection level, Digit 11 – loops to perform, Your hal tester – Seaward ClareHAL User Manual
Page 124

123
Digit 10 – Arc Detection Level
(H101, H103 H104 AC/DC Hipot &
DC IR)
Digit 10 represents a level of arcing detection, which will terminate the
current test sequence when selected level is reached.
Digit 10 : Arc Detection Level
‘0’
Disabled / Off
‘1’
Most sensitive setting – All arcs terminate
‘2’
‘3’
‘4’
‘5’ Medium
sensitivity
‘6’
‘7’
‘8’
‘9’ Least
sensitive
– Only large arcs
Digit 11 – Loops to Perform
Digit 11 represents the number of times a test or linked sequence
should be performed.
The user may use the red STOP key before or during a test and from
the ‘abort options menu’ elect to move to the next test or the next test
after a linked sequence of tests if a smaller number of loops is required.
The STOP button will also terminate an unlimited number of loop tests.
The value of ‘0’ indicates that a test is as ‘follow-on’ test and forms
part of a linked sequence. The linked sequence begins with a test
code that has a non-zero loops value and ends when there are no more
test codes or when the next test code has a non-zero loops value. A
linked sequence is performed the number of times specified by the first
test in the sequence.
Your HAL tester