109
Digit 5 – Ramp to Target Time
Digit 6 – Hold Time
Digit 7 – Ramp to Zero Time
Digits 5, 6 and 7 represent times during which the out voltage is ramped to the target
voltage, held at the target voltage and finally ramped down to zero. All the times are in
seconds.
*Digit 4 is also used in conjunction with Powered Leakage tests to define the test
duration.
Digit 5
Ramp to Target Time
Digit 6 (4)*
Hold Time
Digit 7
Ramp to Zero
Time
‘0’ 0.0
0.0 0.0
‘1’ 0.1
0.1 0.1
‘2’ 0.2
0.2 0.2
‘3’ 0.3
0.3 0.3
‘4’ 0.4
0.4 0.4
‘5’ 0.5
0.5 0.5
‘6’ 0.6
0.6 0.6
‘7’ 0.7
0.7 0.7
‘8’ 0.8
0.8 0.8
‘9’ 0.9
0.9 0.9
‘A’ 1.0
1.0 1.0
‘B’ 1.5
1.5 1.5
‘C’ 2.0
2.0 2.0
‘D’ 3.0
3.0 3.0
‘E’ 4.0
4.0 4.0
‘F’ 5.0
5.0 5.0
‘G’ 7.5
7.5 7.5
‘H’ 10.0
10.0 10.0
‘I’ 15.0
15.0
15.0
‘J’ 20.0
20.0 20.0
‘K’ 30.0
30.0 30.0
‘L’ 45.0
45.0 45.0
‘M’ 60.0
60.0 60.0
‘N’ 90.0
90.0 90.0
‘O’ 120.0
120.0
120.0
‘P’ 150.0
150.0
150.0
‘Q’ 180.0
180.0
180.0
Your HAL tester