Using the Configure Engineering Units Dialog Box

You can use DaqView’s y = mx + b scaling factors by double-clicking in the channel’s Units cell.

This brings up the Configure Engineering Units dialog box (following figure).

Using the Configure Engineering Units Dialog Box

You can use the dialog box to compute the contribution of both 16-bit results to the final 32-bit result so

they appear as engineering units. You can manually add both numbers directly from the screen or later

during post analysis, from data saved in a file.

The appropriate conversions to use are as follows:

counter or totalize mode 32H*65535

32L

period modes

1

[32H*65535]*tickduration/number of periods
[32L]*tickduration/number of periods

pulsewidth

1

or timing

1

modes

[32H*65535]*tickduration
[32L]*tickduration

frequency mode (in Hz)

assuming counter mode

[32H*65535]*scan rate

[

32L]*scan rate

(Note 2)

1

time, e.g., seconds, nanoseconds, microseconds, or milliseconds

2

Where scan rate is expressed in units of: number of scans per second.

Appendix A

907994

Converting Counter Readings to Engineering Units for 3000 Series Devices A-3