Remote Programming
Sorensen SLM-Series DC Load
5-28
M540071-01 Rev C
The dynamic mode's Tlow or Thigh duration query command
Command Syntax:
SLM DC, SLD Modules: PERI:{LOW|HIGH}?{NL}
Description:
PERI:LOW? returns the presently programmed low duration time in dynamic load mode
PERI:HIGH? returns the presently programmed high duration time in dynamic load mode
The engineering unit is “msec”.
RISE Time
Purpose:
The dynamic load mode's RISE slew rate query command
Command Syntax:
SLM DC Modules:
RISE?{NL}
Description:
RISE? returns the presently programmed low load current level in Constant Current mode
The engineering unit is “A/us”.
FALL Time
Purpose:
The dynamic load mode's FALL slew rate query command.
Command Syntax:
SLM DC Modules:
FALL?{NL}
Description:
FALL? returns the presently programmed low load current level in Constant Current mode
The engineering unit is “A/us”.
SHORT ON/OFF
Purpose:
SHORt ON or OFF status query command.
Command Syntax:
All Modules:
SHOR?{NL}
Description:
SHOR? returns the presently SHORt status, “0” indicates SHORt OFF, and “1” indicates SHORt ON.
DYNAMIC ON/OFF
Purpose:
DYNamic ON or OFF status query command
Command Syntax:
SLM DC, SLD Modules: DYN?{NL}
Description:
DYN? returns the presently DYNamic ON or OFF status, “0” indicates static load mode or DYNamic
OFF, and “1” indicates DYNamic load mode or DYNamic ON.