beautypg.com

4 gpib/rs-232 command description, 1 setting commands, Gpib/rs-232 command description – AMETEK SLM Series RevC User Manual

Page 63

background image

Sorensen SLM-Series DC Load

Remote Programming

M540071-01 Rev C

5-7

REMARKS:

1. d : 0 - 9
2. GLOB : GLOBAL (ALL CHANNELS ACTIVE AT SAME TIME)
3. CURRENT ENGINEERING UNIT : A
4. VOLTAGE ENGINEERING UNIT : V
5. RESISTANCE ENGINEERING UNIT :

Ω

6. PERIOD ENGINEERING UNIT : mS
7. SLEW-RATE ENGINEERING UNIT : A/

μS


Note

: The RS-232 command set is the same as the GPIB command set.

5.4

GPIB/RS-232 Command Description

5.4.1 Setting

Commands

CHANNEL

Purpose:
”CHAN" selects the multiple Electronic load channel to which all subsequent channel-specific
commands will be directed.

Command Syntax:
All Modules:

CHAN{SP}n{;NL}

Description:
”CHAN" command selects the specified Electronic load module from 1 through 4 as the Electronic
load module number (from left to right). Up to 4 channels of the Electronic load module can be
installed in one mainframe. This command is a channel independent command; therefore, this
command should be programmed before an electronic load channel dependent command.
The load channel number is arranged as 1, 2, 3, 4 from left side to the right side. Module SLD-60-
105-550 is a dual bay module that utilizes the channel number corresponding to the bay occupied
by the right side of the module. E.g., if a dual bay module takes up bays 1 and 2, its channel
number shall be 2.

Example:

CHAN 2 select channel 2 of SL-series mainframe.

Note:

Please refer to Appendices C, D and E for proper programming procedure of SL-series electronic load

modules.

CURRENT

Level

Purpose:
The load current setting in Constant Current mode.

Command Syntax:
All SLM Modules:

CC:{LOW|HIGH}{SP}{NR2}{;|NL}

SLD Modules:

CC{SP}{NR2}{;|NL}