AMETEK SLM Series RevC User Manual

Page 45

background image

Sorensen SLM-Series DC Load

Operation

M540071-01 Rev C

3-17

value via GPIB, RS-232, or the front panel. Figure 3-6 shows the analog programming signal
(4 Vac, 500Hz) summed with the 24A programmed setting in CC mode.

2mS

T

LOAD CURRENT

6V

4V

2V

12A

24A

36A

ANALOG V

Figure 3-6 Analog Programming Load Current in CC Mode Operation


Caution:

The analog input is always active. Even if the load is in the “OFF”

state at the front panel or through remote control, the load will draw current when
voltage is applied to the analog input.

3.2.6

Load Coarse/Fine Increase/Decrease Adjustment Key

The load current step change or resolution is shown in Table 3-2 for each load module.

Pressing and holding any one of the four arrow keys for more than 1 second speeds the step change;
either release the arrow key to stop at a value, or wait for the adjustment to reach the maximum or
minimum load current setting.

SLM-60-15-75

RANGE 1

RANGE 2

COARSE/FINE LOAD
CURRENT ADJUSTMENT
KEY

×Ø

↓↑

×Ø

↓↑

CC Mode

4.0 mA

0.4 mA

40.0 mA

4.0 mA

CR Mode

6.666mS

0.0666mS

0.1066

Ω 1.066mΩ

CV

Mode

0.16V 16mV 0.16V 16mV

CP

Mode

0.2W 0.02W 0.2W 0.02W


SLM-60-30-150

RANGE 1

RANGE 2

COARSE/FINE LOAD
CURRENT ADJUSTMENT
KEY

×Ø

↓↑

×Ø

↓↑

CC Mode

8.0 mA

0.8 mA

80 mA

8.0 mA

CR Mode

13.33mS

0.133mS

53.33m

Ω 0.533mΩ

CV

Mode

0.16V 0.016V 0.16V 0.016V

CP

Mode

0.4W 0.04W 0.4W 0.04W


(Continued)