AMETEK SLM Series RevC User Manual
Page 45

Sorensen SLM-Series DC Load
Operation
M540071-01 Rev C
3-17
value via GPIB, RS-232, or the front panel. Figure 3-6 shows the analog programming signal
(4 Vac, 500Hz) summed with the 24A programmed setting in CC mode.
2mS
T
LOAD CURRENT
6V
4V
2V
12A
24A
36A
ANALOG V
Figure 3-6 Analog Programming Load Current in CC Mode Operation
Caution:
The analog input is always active. Even if the load is in the “OFF”
state at the front panel or through remote control, the load will draw current when
voltage is applied to the analog input.
3.2.6
Load Coarse/Fine Increase/Decrease Adjustment Key
The load current step change or resolution is shown in Table 3-2 for each load module.
Pressing and holding any one of the four arrow keys for more than 1 second speeds the step change;
either release the arrow key to stop at a value, or wait for the adjustment to reach the maximum or
minimum load current setting.
SLM-60-15-75
RANGE 1
RANGE 2
COARSE/FINE LOAD
CURRENT ADJUSTMENT
KEY
×Ø
↓↑
×Ø
↓↑
CC Mode
4.0 mA
0.4 mA
40.0 mA
4.0 mA
CR Mode
6.666mS
0.0666mS
0.1066
Ω 1.066mΩ
CV
Mode
0.16V 16mV 0.16V 16mV
CP
Mode
0.2W 0.02W 0.2W 0.02W
SLM-60-30-150
RANGE 1
RANGE 2
COARSE/FINE LOAD
CURRENT ADJUSTMENT
KEY
×Ø
↓↑
×Ø
↓↑
CC Mode
8.0 mA
0.8 mA
80 mA
8.0 mA
CR Mode
13.33mS
0.133mS
53.33m
Ω 0.533mΩ
CV
Mode
0.16V 0.016V 0.16V 0.016V
CP
Mode
0.4W 0.04W 0.4W 0.04W
(Continued)