Operation

Sorensen SLM-Series DC Load

3-16

M540071-01 Rev C

3.2.4

Initial Setting of SLM-Series DC Load Modules

Table 3-1 lists the factory, or initial, load value settings of the SLM-series DC electronic load when it is
initially powered ON once it first arrives from the factory.

Item

SLM-60-15-75 SLM-60-30-150 SLM-60-60-300 SLM-250-10-300 SLM-500-10-300

Basic

Settings

CC LOW [amps]

0.0000

0.0000

0.000

0.0000

0.0000

CC HIGH [amps]

0.0000

0.0000

0.000

0.0000

0.0000

CR HIGH [

Ω] 15000. 7500. 3750. 18750. 18750.

CR LOW [

Ω] 15000. 7500. 3750. 18750.

18750.

CV HIGH [volts]

60.00

60.00

00.00

250.0

500.0

CV LOW [volts]

60.00

60.00

60.00

250.0

500.0

CP LOW [watts]

0.000

0.000

0.000

0.000

0.000

CP HIGH [watts]

0.000

0.000

0.000

0.000

0.000

Dynamic Settings

FALL [A/

μs] 0.0625 0.1250 0.2500 0.0500 0.0250

RISE [A/

μs] 0.0625 0.1250 0.2500 0.0500 0.0250

T

HIGH

[ms]

0.0500 0.0500 0.0500 0.0500 0.0500

T

LOW

[ms]

0.0500 0.0500 0.0500 0.0500 0.0500

Go/No Go Limits

V

HIGH

60.00 60.00 60.00 300.0 600.0

V

LOW

0.000 0.000 0.000 0.000 0.000

CURR.

HIGH 15.000 30.00 60.00 12.000 6.000

CURRENT

LOW

0.000 0.000 0.000 0.000 0.000

POWER

HIGH 75.00 150.0 300.0 300.0 200.0

POWER

LOW 0.000 0.000 0.000 0.000 0.000

System Settings

V

SENSE

Auto Auto Auto Auto Auto

LD-ON

[volts] 1.000 1.000 1.000 2.000 4.000

LD-OFF [volts]

0.5

0.5

0.5

0.150

0.300

Table 3-1 SLM-Series DC Electronic Load Factory Settings for Power On

3.2.5

Analog Programming Input

The Electronic Load has an analog programming 9-pin input on the rear panel of SLM
mainframe, labeled CH1, CH2, CH3, and CH4 respectively. See the SLM Mainframe
Operations Manual for connection instructions. CC and CP modes can be programmed with
a 0 to 10V analog signal (arbitrary waveform or arbitrary waveform + dc) connected to the
analog programming input.

Example: SLM-60-60-300. The 0 to 10V Analog signal can program the 0 to
full-scale input Range 1n the CC mode (0 to 6A range when load current setting
is less than 6A, or 0 to 60A range when load current setting is higher than 6A) or
in the CP mode (0 to 300W). To be operated only in the high range (Range 2).

The analog programming signal can act alone or it can be summed with the programmed